Navios EX Flow Cytometer

The Navios EX flow cytometer offers a solution for advanced cytometry applications with workflows for high throughput laboratories. This is the next generation leveraging the laser and optics technology of our successful CytoFLEX research platform. Robust performance means peace of mind, because when analyzing important samples, every event matters.

  • Up to 12 parameters for high complexity assays*; sensitivity and resolution; and quality, reproducible results
  • With 10 colors*, collect additional data points from each sample, 
  • Reduce the number of samples to prepare along with the possible pipetting errors
 

Choose a Model

Part Number Product Name Status List Price Buy

Optical System

  • Up to three lasers, 488 nm blue laser, 638 nm red laser and 405 nm violet laser
  • Solid state independently focused diode lasers
  • Integrated flow cell minimizes light loss

Clinical Workflow

  • Task scheduler powers up or off
  • Bi-directionally interface with Laboratory Information Systems
  • Manage test orders, create work lists, and track samples

Flexible Detection

  • Wide angle scatter for distinguishing cells from debris
  • Narrow angle scatter for detecting larger particles
  • Scatter signal amplification for submicron particle detection

Stable Performance

  • Thermoregulation system maintains the optical area at a consistent temperature
  • Compensation settings maintained over time
  • Accurate digital signal processing at high event rates and over time

Product Specifications

Detection Photodiodes (forward & side scatter), Photomultiplier tubes (fluorescence)
Resolution HPCV is <2%
Performance <107 MESF for FITC, <64 MESF for PE, <13 MESF for PC5
Flow Cell 430 μm x 180 μm rectangular channel
Signal Processing Digital System with 4 decade data display
Power Requirements 100-240V, 50-60Hz

Documentation

  • Filter by:
Title Description
*NAVIOS EX is CE marked for 10-color in-vitro diagnostic use. In the US, the Navios EX Flow Cytometer is intended for use as an in vitro diagnostic device for immunophenotyping using up to four fluorescent detection channels using a blue (488 nm) laser and two light scatter detection channels. It is intended for use with in vitro diagnostic (IVD) assays and software that are indicated for use with the instrument.

Technical Documents