MET ONE R4800 and R4900
The MET ONE R4800 and R4900 series remote air particle counters are affordable, reliable and provide accurate data in particle monitoring systems. A compact design with flexible sizing, flow rate and communication options address the specific needs of cleanroom operations in pharmaceutical, semiconductor, hard disk drive and flat panel display industries.
Choose a MET ONE R4800 or R4900 Model
Tech Docs | Part Number | Product Name | Description | Flow Rate | Particle Size Analysis Range | List Price | Buy |
---|---|---|---|---|---|---|---|
|
|
0.5 - 5 μm / 0.5 - 5 μm
|
- |
||||
|
|
0.5 - 5 μm / 0.5 - 5 μm
|
- |
||||
|
|
0.3 - 5 μm / 0.3 - 5 μm
|
- |
||||
|
|
0.5 - 5 μm / 0.5 - 5 μm
|
- |
||||
|
|
0.5 - 5 μm / 0.5 - 5 μm
|
- |
||||
|
|
0.3 - 0 μm / 0.3 - 0 μm
|
- |
||||
|
|
0.5 - 5 μm / 0.5 - 5 μm
|
- |
||||
|
|
0.3 - 0 μm / 0.3 - 0 μm
|
- |
Features
Easy Integration
- MET ONE R4800 series remote air particle counters use standard RS-485 serial communications
- MET ONE R4900 series remote air particle counters use industry standard 4-20mA interface
- Stainless steel enclosure compatible with most process environments
- Compact size for easy mounting with a vacuum source, DC power & communication line
Data Management
- MET ONE R4800 & R4900 counters respond to simple ASCII command instructions, including start, stop and report counts, send location number, etc.
- Reporting data includes counter status & particle counts in two size ranges
- Multiple display options via continuous monitoring feature (Particle Vision® Online & CIMScan® software
Operation
- User-programmed alarm level detects high counts on first channel triggering an alarm
- Auto alarm reset at start of the next count cycle
- Long-life laser technology for superior performance & extended service life
- No fan allows proximate placement to sensitive processing areas
Flexible
- Sensitivity ranges of 0.5um & flow rates of 1.0 cfm
- Onboard status indicator sets signal at connector for alarm Stay resident RS-485 configuration programming (count/hold times, alarm level/delay & scaling) maintained through power cycles
- Sensor resistant to vaporous hydrogen peroxide (VHP)
Product Specifications
Particle Size Analysis Range | 0.50|5.00|μm / 0.5|5|μm |
Coincidence Loss | 5% at 70,630,000 particles/m3 |
Throughput | 28.32 L per minute |
Count Alarms Display and Cycles | User-defined alarm counts displayed via LED on box or remote LED |
Item Specifications Referenced | 2086895-04 |