CytoFLEX nano級流式細胞儀

CytoFLEX nano 流式細胞分析儀是 CytoFLEX 平台的最新成員。.

CytoFLEX nano 流式細胞儀專為分析奈米粒子而設計,它克服了傳統流式細胞儀的限制,為研究人員提供了靈敏, 一致靈活的解決方案,有助於他們推進研究。

CytoFLEX nano 流式細胞儀為流式細胞儀學的重要發展,可讓使用者輕鬆分析至少小至 40 nm* 的細胞外小泡 (EV),同時提供多達 6 個獨立的螢光檢測通道及 5 個側向散射通道,以提供完整的特性分析,並為您的奈米粒子研究開闢更多的途徑。

CytoFLEX nano Flow Cytometer Features

Explore the full picture of your EV experiment with the CytoFLEX nano Flow Cytometer. This purpose-built flow cytometer provides greater sensitivity, consistent instrument performance, and flexibility to study your sample.

靈敏度

CytoFLEX nano 流式細胞儀具有從 40nm*到 1µm 的無與倫比的尺寸靈敏度和 10nm 的解析度**,能夠檢測較小的 EV 群及其低豐度的貨物。

靈活性

利用 6 個螢光通道和 5 個側向散射通道開啟您的研究。以更少的限制收集所有必要的資料。

最大一致性

在處理小尺寸顆粒時,維持無污染、穩定的儀器效能至關重要。透過自動 QC 控制,體積計數準確度 >90%,樣品間攜帶率 <1%,可獲得值得信賴的再現結果。

熟悉性

CytoFLEX nano 流式細胞儀基於與 CytoFLEX 平台相同的原理製造而成,具有相似的體積與友善的軟體介面。

 



The CytoFLEX nano analyzer utilizes  CytExpert for CytoFLEX Acquisition and Analysis Software and has the ability to count, size, and characterize EVs using a single technique, propelling your research forward.

Size  Count  Characterize
Particle size determination by flow cytometry requires calibration using an appropriately characterized reference particle paired with a calibration method. Latex polystyrene particles are the tool of choice for this task due to their NIST-traceable nature and reliably consistent manufacturing. Through the calibration method taught to each CytoFLEX nano flow cytometer customer, data will be reliably reportable in absolute units for method comparison and correlation. The most robust counting methods rely on highly controlled volumetric delivery of samples rather than hand loading samples with the implied variability. The CytoFLEX nano flow cytometer allows users to customize their sample handling so that the right amount of sample is used for the results needed. The ability to describe nanoscale particles requires best-in-class sensitivity. To make this possible, the CytoFLEX nano flow cytometer has completely redesigned the optics and fluidics system to maximize the detection of even the smallest amounts of fluorescence across 6 detection channels and excited by 4 collinear lasers.

性能特性

CytoFLEX nano Flow Cytometer Performance
 Violet side scatter sensitivity VSSC1: 40 nm relative to polystyrene nanoparticles
VSSC2: 80 nm relative to polystyrene nanoparticles
Scatter detection dynamic range VSSC 1 for small range 40-150 nm polystyrene nanoparticles
VSSC2 for large range 80-1000 nm polystyrene nanoparticles
Violet forward scatter sensitivity 300 nm relative to polystyrene nanoparticles
Fluorescence sensitivity and resolution

 Simultaneous detection of fluorescence on six fluorescence detectors, using 500 nm CytoFLEX nano Multi-intensity Fluorospheres

V447: 8 peaks

B531: 8 peaks

Y595: 8 peaks

R670: 6 peaks

R710: 5 peaks

R792: 4 peaks

 Fluorescence rCV CV < 10% (using QC Fluorospheres at 1 µL/min)
Carryover between Samples 1%a
 Acquisition speed Maximum electronic acquisition speed 16000 events/second with ≤ 95% yield
Recommended maximum sample acquisition speed 5000 events/second to avoid possible swarming or coincidence situations.
 Volumetric Counting accuracy > 90%b

Learn more about nanoscale flow cytometry

Explore CytoFLEX nano Models

a. Tested with polystyrene nanoparticles

b. Tested with 144 nm QC Scatterspheres on 3 units under the conditions: record 3 minutes at the sample flow rates of 1µL/min, 2µL/min, 3µL/min, 4µL/min, 5µL/min, 6µL/min respectively, repeat 5 times, then calculate the average total events for each speed, then compare the calculated volumetric counting accuracy with the theoretical total events ( the known concentration 1.3*10 8 x volume).

*polystyrene when triggering on violet side scatter
** In development. Performance characteristics have not been validated.

For Research Use Only. Not for use in diagnostic procedures.